Abstract
In spite of substantial technological advances in analytical field, most instruments cannot directly handle complex sample matrixes yet. As a result, a sample-preparation step is commonly involved before instrumental analysis. In the proposed study, a new dispersive liquid-liquid micro extraction (DLLME) method based on surfactant was developed for the determination of Aluminum (Al) in tap water. The enrichment factor, limit of detection and relative standard deviation of the suggested method were found to be 10, 0.023 mg / L (n = 5, 3s / b), 97 ± 1.6, respectively. The affecting parameters on the optimal determination conditions such as pH, suitable extracting solvent and disperser solvent, volume of extracting solvent and disperser solvent and affects of diverse ions, were investigated. The method was successfully applied to drinking water. In order to test the accuracy of the proposed method, recovery studies were conducted on tap...